Your bispecific gained an acidic peak. Does it still bind both targets?

Say a new manufacturing lot comes back with a larger acidic peak. The unfractionated antibody still binds about as well as the reference. Would you move on?

I’d collect that peak and test both arms separately. In a Bio-Techne and Nicoya application note on the CD20 × CD3 bispecific mosunetuzumab, two unfractionated samples had nearly identical CD20 affinities. Yet within the research-grade comparison material, the acidic fraction had a 7.2-fold higher apparent CD20 K_D than its own main fraction. Its CD3 affinity was close to the main fraction’s. If I’d only run the bulk sample or checked the CD3 arm, I would have missed the difference.

Why Alto is a good match for MauriceFlex

Bio-Techne’s MauriceFlex uses imaged capillary isoelectric focusing (icIEF) to separate proteins by their isoelectric point, or pI. That’s the pH at which a protein has no net charge. An acidic peak focuses at a lower pH than the main peak. MauriceFlex can collect the material behind those peaks in a 96-well plate, so you can take the acidic and main fractions into the next assay.

The two systems work with small volumes at different points in the workflow. Bio-Techne specifies 100 µL of prepared sample as the minimum for MauriceFlex fractionation. Once you’ve collected a fraction, Alto Digital SPR uses surface plasmon resonance (SPR) to follow how quickly it binds and how quickly the complex comes apart. In the mosunetuzumab study, the team loaded 2 µL of each prepared antibody fraction for the CD3 assay. A 2 µL load is useful when I want binding curves from a fraction and still need material for a repeat or mass spec. The authors pooled and concentrated fractions from four runs per material, so I’d check recovery and concentration before allocating the sample.

Once the fractions are prepared, the transfer is a pipetting step. Load an aliquot onto the disposable Alto cartridge, and Alto handles the serial dilutions. For CD3, it made five concentrations of the antibody fraction. For CD20, the team captured the antibody and Alto made five concentrations of the antigen. That saves a round of manual dilution for each series. The liquid handling is contained in the cartridge, which you remove at the end of the run.

I can picture the pairing on a shared bench, too. MauriceFlex is 42 cm wide by 61 cm deep, and Alto is 35 cm wide by 52 cm deep in Nicoya’s technical brochure. The two instrument bodies add up to about 77 cm (30 in) of width side by side. I’d allow extra room for the workstation, ventilation, and access to both instruments.

The CD20 difference only showed up after fractionation

These are the reported apparent dissociation constants, or K_D values. Within the same assay, a higher K_D indicates weaker apparent binding. The authors describe the comparison material as a research-grade biosimilar.

Sample CD3 epsilon K_D CD20 nanodisc K_D
Reference mosunetuzumab, unfractionated 28.6 nM 12.9 nM
Research-grade comparison, unfractionated 22.8 nM 13.4 nM
Research-grade comparison, main fraction 14.7 nM 22.4 nM
Research-grade comparison, acidic fraction 12.9 nM 162 nM

The assay setup mattered. The team immobilized CD3 epsilon on the sensor and flowed the antibody fractions over it. For CD20, they captured the antibodies by their Fc region and flowed full-length CD20 carried in a nanodisc, a tiny patch of lipid membrane. The recombinant CD20 extracellular construct they tried gave no appreciable binding in their screen. If an arm looks weak, I’d check how the antigen was presented before blaming the antibody.

What I’d do with a growing acidic peak

  1. Collect the acidic and main fractions and keep a bulk control. I’d confirm that the fractions still map to the expected peaks and check how much antibody I recovered before splitting the material between assays.
  2. Measure each arm and look at the curves. I’d compare the fractions within each target assay. For CD20, I’d check how much antibody each sensor captured. I’d also look at the reference and blank signals, and repeat a meaningful difference with independently collected material. For more on reading the curves, see how much can you trust the off-rate?
  3. Match the next test to the decision. For molecular identity, I’d send an aliquot to liquid chromatography–mass spectrometry (LC–MS). If I need to know whether the same molecule can engage both targets at once, I’d use a bridging assay designed for that question. If the decision is about potency, I’d use a cell-based assay tied to how the antibody works. The two separate SPR assays in this note establish binding to each target at the fraction level.

The mass spec follow-up is especially interesting here. In a companion LC–MS study, Bio-Techne and Genovis found correctly paired antibodies and antibodies with mismatched light chains in the acidic comparison fractions. That gives you a lead to investigate. The 162 nM value describes a mixed fraction, so it can’t tell you the affinity of the mismatched species on its own.

An acidic peak can also turn out to have little effect in the assays you care about. Wu and colleagues reported a bispecific with an unexpected disulfide bond and comparable binding and potency in its acidic and main peaks. Nguyen and colleagues found reduced cell-based potency in some acidic fractions of other anti-CD3 bispecifics. I touched on the limits of a reassuring bulk result in my piece on what changed when you added the payload. Here, testing the fractions showed exactly which interaction deserved a closer look.

Give the peak a follow-up experiment

If I already had MauriceFlex, I’d pair it with Alto so I could test both arms with small aliquots and keep material for whichever interaction needs a closer look. In this case, that would be CD20. If both arms still bind separately and the molecule is designed to bring two cells together, I’d test whether it can still do that. That’s the result I’d want in hand before deciding whether the peak belongs on a trend chart or at the top of the process investigation.